Sensitive Detection of Black Powder by a Stand-Alone Ion Mobility Spectrometer with an Embedded Titration Region
Liang, Xixi1,2; Zhou, Qinghua1,2; Wang, Weiguo1; Wang, Xin1; Chen, Wendong1,2; Chen, Chuang1,2; Li, Yang1; Hou, Keyong1; Li, Jinghua1; Li, Haiyang1
刊名analytical chemistry
2013-05-21
卷号85期号:10页码:4849-4852
通讯作者李海洋
产权排序待补充
合作状况
英文摘要sensitive detection of black powder (bp) by stand-alone ion mobility spectrometry (ims) is full of challenges. in conventional air-based ims, overlap between the reactant ion o-2(-)(h2o)(n) peak and the sulfur ion peak occurs severely; and common doping methods, providing alternative reactant ion cl-(h2o)(n), would hinder the formation of ionic sulfur allotropes. in this work, an ion mobility spectrometer embedded with a titration region (tr-ims) downstream from the ionization region was developed for selective and sensitive detection of sulfur in bp with ch2cl2 as the titration reagent. sulfur. ions were produced via reactions between sulfur molecules and o-2(-)(h2o)(n) ions in the ionization region, and the remaining o-2(-)(h2o)(n) ions that entered the titration region were converted to cl-(h2o)(n) ions, which avoided the peak overlap as well as the negative effect of ch2cl2 on sulfur ions. the limit of detection for sulfur was measured to be 5 pg. furthermore, it was demonstrated that this tr-ims was qualified for detecting less than 5 ng of bp and other nitro-organic explosives.
学科主题物理化学
WOS标题词science & technology ; physical sciences
类目[WOS]chemistry, analytical
研究领域[WOS]chemistry
关键词[WOS]explosive trace detector ; mass-spectrometer ; sulfur
收录类别SCI
资助信息1,1
原文出处4852
语种英语
WOS记录号WOS:000319651100004
公开日期2014-09-11
内容类型期刊论文
源URL[http://159.226.238.44/handle/321008/119305]  
专题大连化学物理研究所_中国科学院大连化学物理研究所
作者单位1.Chinese Acad Sci, Dalian Inst Chem Phys, Dalian 116023, Peoples R China
2.Chinese Acad Sci, Grad Univ, Beijing 100039, Peoples R China
推荐引用方式
GB/T 7714
Liang, Xixi,Zhou, Qinghua,Wang, Weiguo,et al. Sensitive Detection of Black Powder by a Stand-Alone Ion Mobility Spectrometer with an Embedded Titration Region[J]. analytical chemistry,2013,85(10):4849-4852.
APA Liang, Xixi.,Zhou, Qinghua.,Wang, Weiguo.,Wang, Xin.,Chen, Wendong.,...&Li, Haiyang.(2013).Sensitive Detection of Black Powder by a Stand-Alone Ion Mobility Spectrometer with an Embedded Titration Region.analytical chemistry,85(10),4849-4852.
MLA Liang, Xixi,et al."Sensitive Detection of Black Powder by a Stand-Alone Ion Mobility Spectrometer with an Embedded Titration Region".analytical chemistry 85.10(2013):4849-4852.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace