英文摘要 | The robotic-assisted percutaneous coronary intervention is an emerging technology with great potential tosolve the shortcomings of existing treatments. However, thecurrent robotic systems can not manipulate two guidewires or ballons/stents simultaneously for coronary bifurcation lesions.This paper presents VasCure, a novel bio-inspired vascular robotic system, to deliver two guidewires and stents into themain branch and side branch of bifurcation lesions in sequence. The system is designed in master-slave architecture to reduce occupational hazards of radiation exposure and orthopedic injury to interventional surgeons. The slave delivery device has one active roller and two passive rollers to manipulate two interventional devices. The performance of the VasCure was verified by in vitro and in vivo animal experiments. In vitro results showed the robotic system has good accuracy to deliverguidewires and the maximum error is 0.38mm. In an animal experiment, the interventional surgeon delivered two guidewires and balloons to the left circumflex branch and the left anterior descending branch of the pig, which confirmed the feasibility of the vascular robotic system. |
会议录出版者 | IEEE |
语种 | 中文 |
内容类型 | 会议论文 |
源URL | [http://ir.ia.ac.cn/handle/173211/48705] |
专题 | 自动化研究所_复杂系统管理与控制国家重点实验室_先进机器人控制团队 |
通讯作者 | Xiao-Liang Xie |
作者单位 | 1.School of Artificial Intelligence, University of Chinese Academy of Sciences 2.CAS Center for Excellence in Brain Science and Intelligence Technology 3.China University of Mining and Technology 4.State Key Laboratory of Management and Control for Complex Systems, Institute of Automation, Chinese Academy of Sciences |
推荐引用方式 GB/T 7714 | Han-Lin Zhao,Shi-Qi Liu,Xiao-Hu Zhou,,et al. Design and Performance Evaluation of a Novel Vascular Robotic System for Complex Percutaneous Coronary Interventions[C]. 见:. Mexico. 2021.11.1 - 2021.11.5. |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论