Isochronicity corrections for isochronous mass measurements at the HIRFL-CSRe | |
Yin, Da-Yu2; Shang, Peng1,2; Shi, Jian2; Chai, Wei-ping2; Shen, Guo-Dong1,2; Zhang, Wei2; Litvinov, Y.3; Wang, Meng2; Litvinov, S.3; Yang, Jian-cheng2 | |
刊名 | NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT |
2014-11-01 | |
卷号 | 763页码:53-57 |
关键词 | Isochronous Mass Measurement Sextupoles Correction Second-order Isochronicity |
ISSN号 | 0168-9002 |
DOI | 10.1016/j.nima.2014.05.122 |
文献子类 | Article |
英文摘要 | Isochronous Mass Spectrometry (IMS) is a unique experimental method for mass measurement experiments on short-lived nuclei. Mass measurements of Kr-78 projectile fragments were performed in HIRFL-CSRe at the Institute of Modern Physics (IMP), Chinese Academy of Sciences. The short-lived secondary beams were produced by bombarding a 15 mm thick beryllium target in the Radioactive Ion Beam Line (RIBLL2) and were then injected into the CSRe storage ring. The masses of stored ions were measured by employing the IMS technique, which is based on the determination of the ion revolution Limes. A dedicated time-of-flight. (TOP) defector is used for the latter purpose. However, the isochronicity, and thus the mass resolving power, depends on the momentum spread and the transverse ernittance of the injected beams. Here, we present. the first order isochronicity optimization, the chromaticity and second order isochronicity corrections through the modification of the quadrupole and sextupole field strengths. With the help of these corrections, the mass resolution of Delta m/m = 10(-6) can be achieved. (C) 2014 Elsevier B.V. All rights reserved. |
WOS关键词 | RING PROJECT ; BEAM |
WOS研究方向 | Instruments & Instrumentation ; Nuclear Science & Technology ; Physics |
语种 | 英语 |
出版者 | ELSEVIER SCIENCE BV |
WOS记录号 | WOS:000341986200008 |
内容类型 | 期刊论文 |
源URL | [http://119.78.100.186/handle/113462/49721] |
专题 | 近代物理研究所_实验物理中心 近代物理研究所_加速器总体室 |
通讯作者 | Yang, Jian-cheng |
作者单位 | 1.Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China 2.Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China 3.Helmholtzzentrum Schwerionenforsch GmbH, GSI, D-64291 Darmstadt, Germany |
推荐引用方式 GB/T 7714 | Yin, Da-Yu,Shang, Peng,Shi, Jian,et al. Isochronicity corrections for isochronous mass measurements at the HIRFL-CSRe[J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT,2014,763:53-57. |
APA | Yin, Da-Yu.,Shang, Peng.,Shi, Jian.,Chai, Wei-ping.,Shen, Guo-Dong.,...&Gao, Xiang.(2014).Isochronicity corrections for isochronous mass measurements at the HIRFL-CSRe.NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT,763,53-57. |
MLA | Yin, Da-Yu,et al."Isochronicity corrections for isochronous mass measurements at the HIRFL-CSRe".NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT 763(2014):53-57. |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论