TL response of LiF:Mg,Cu,P (GR200A and GR207A) exposed to high-energy C-12 ions | |
Ma, F.![]() ![]() ![]() | |
刊名 | RADIATION MEASUREMENTS
![]() |
2015-07-01 | |
卷号 | 78页码:23-27 |
关键词 | Thermoluminescent detectors GR200A and GR207A High-energy C-12 ions Dose response Thermoluminescent efficiency |
ISSN号 | 1350-4487 |
DOI | 10.1016/j.radmeas.2014.10.004 |
英文摘要 | The characteristics of (LiF)-Li-nat.,7:Mg,Cu,P thermoluminescent detectors exposed to high-energy C-12 (165 MeV/nucleon) ions and Co-66 gamma-rays have been investigated. Through a careful analysis of the glow curves by the GlowFit program, it is found that the relative amplitude of satellite peaks 3, 5, 6 and 7 to peak 4, measured after irradiated with high-energy C-12 ions, is higher than that with Co-60 gamma-rays for both types of TL, detectors. It is observed that "(LiF)-Li-nat.,7:Mg,Cu,P thermoluminescent detectors would also exhibit a good linear dose response when the delivered dose of high-energy C-12 ions is in the range from 100 mGy to 1600 mGy. Furthermore, the relative efficiency of the (LiF)-Li-nat.,7:Mg,Cu,P thermoluminescent detectors is independent of the absorbed dose in the linear dose response range. Compared to previously published data, the relative efficiency of the (LiF)-Li-7:Mg,Cu,P TL detectors decreases with increasing LET values. (C) 2014 Elsevier Ltd. All rights reserved. |
资助项目 | National Natural Science Foundation of China[91226107] ; National Natural Science Foundation of China[11305229] ; National Natural Science Foundation of China[11105186] ; Strategic Priority Research Program of the Chinese Academy of Sciences[XDA03030300] |
WOS关键词 | TRACK STRUCTURE-THEORY ; THERMOLUMINESCENT DETECTORS ; RELATIVE EFFICIENCY ; LIF-MG,CU,P ; RADIATION ; LIF ; DOSIMETRY ; PARTICLE ; BEAMS |
WOS研究方向 | Nuclear Science & Technology |
语种 | 英语 |
出版者 | PERGAMON-ELSEVIER SCIENCE LTD |
WOS记录号 | WOS:000356747500004 |
资助机构 | National Natural Science Foundation of China ; Strategic Priority Research Program of the Chinese Academy of Sciences |
内容类型 | 期刊论文 |
源URL | [http://119.78.100.186/handle/113462/40054] ![]() |
专题 | 近代物理研究所_实验物理中心 |
通讯作者 | Chen, L. |
作者单位 | Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China |
推荐引用方式 GB/T 7714 | Ma, F.,Ge, H. L.,Zhang, Y. B.,et al. TL response of LiF:Mg,Cu,P (GR200A and GR207A) exposed to high-energy C-12 ions[J]. RADIATION MEASUREMENTS,2015,78:23-27. |
APA | Ma, F..,Ge, H. L..,Zhang, Y. B..,Wan, B..,Li, Y. Y..,...&Zhang, H. B..(2015).TL response of LiF:Mg,Cu,P (GR200A and GR207A) exposed to high-energy C-12 ions.RADIATION MEASUREMENTS,78,23-27. |
MLA | Ma, F.,et al."TL response of LiF:Mg,Cu,P (GR200A and GR207A) exposed to high-energy C-12 ions".RADIATION MEASUREMENTS 78(2015):23-27. |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论