Performance test of PPAC in different gases
Wang, M; Guo, ZY; Zhan, WL; Xiao, GQ; Xu, HS; Mao, RS; Hu, ZG; Chen, ZQ; Sun, ZY; Li, JX
刊名HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION
2002-12-01
卷号26期号:12页码:1285-1290
关键词Parallel-plate Avalanche Counter Isobutane Position Resolution
ISSN号0254-3052
英文摘要

A two-dimension position sensitive parallel-plate avalanche (PPAC) detector has been developed for RIBLL. The detector consists of one anode and two cathodes. In each cathode a resistance chain is used to readout position signals.. The detector has been tested in different operating gases with an a source. When the detector is at 7mb flowing rate of isobutane and + 500V on anode, the position resolution of 0.76mm is obtained. For 7mb C3F8 and + 595V on anode, the position resolution is 0.64mm. The efficiencies are around 99.1% in the cases of C3F8 and isobutane.

WOS研究方向Physics
语种英语
出版者SCIENCE CHINA PRESS
WOS记录号WOS:000180122000013
公开日期2010-10-29
内容类型期刊论文
源URL[http://ir.imp.cas.cn/handle/113462/1569]  
专题近代物理研究所_近代物理研究所知识存储(2010之前)
近代物理研究所_实验物理中心
近代物理研究所_兰州重离子研究装置
中国科学院近代物理研究所
作者单位Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China
推荐引用方式
GB/T 7714
Wang, M,Guo, ZY,Zhan, WL,et al. Performance test of PPAC in different gases[J]. HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION,2002,26(12):1285-1290.
APA Wang, M.,Guo, ZY.,Zhan, WL.,Xiao, GQ.,Xu, HS.,...&Li, CF.(2002).Performance test of PPAC in different gases.HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION,26(12),1285-1290.
MLA Wang, M,et al."Performance test of PPAC in different gases".HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION 26.12(2002):1285-1290.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace