用于中国散裂中子源多功能反射谱仪的高气压多丝正比室探测器的研制
刘凌; Sun, ZJ; Li, YH; Zhang, J; Wang, YF; Liu, L; Wang, HY; Zhang, YL; Qi, HR; Wen, ZW
刊名物理学报
2018
卷号67期号:7页码:72901
关键词multi-wire proportional chamber detection efficiency position resolution n/gamma resolution
ISSN号1000-3290
DOI10.7498/aps.67.20172618
其他题名Development of high-pressure multi-wire proportional chamber neutron detector for the China Spallation Neutron Source multipurpose reflectometer
文献子类Article
英文摘要As a multipurpose reflectometer device, the two-dimensional (2D) position resolution neutron detector with a 200 mm x 200 mm effective area is developed for China Spallation Neutron Source (CSNS) in Dongguang, China. Due to the requirements for the specific parameters of the multipurpose reflectometer, it should be designed to have a more than 50% (@2 angstrom) detection efficiency, better than 2 mm position resolution and 3 times n/gamma resolution ability during the whole operation period of 10 years. The high pressure multi-wire proportional chamber (MWPC) neutron detector filling 3 He gas is used as a key detector. Some simulation results and the experimental results show that the optimized thickness of the neutron entrance window should be 9 mm with using the 7075 aluminum alloy, the high pressure chamber should be sealed by the aluminous ring and a gas mixture should be filled with 6 bar He-3+2.5 bar C3 H8. The assembled detector can achieve a more than 54% (@2 angstrom) detection efficiency in the normal operation. With the 100 mu m wide collimator slit, the position resolution for X-rays is about 0.235 mm. Therefore, the position resolution for neutron is about 1.4 mm when 2.5 bar propane is used as the stopping gas for proton and triton. In the chamber, the water vapor, the oxygen and the organic impurity gases will reduce the gas gain, cause the detector electrodes to break down and the detector to speed up aging. To solve the outgassing effect of the detector components and keep the stable operation, the recycled device is designed to have the purification function for the working gases. It could purify the working gas at a flow rate of 2L/min to remove the oxygen, the water vapor and the organic impurity gases. The detector gain increases about 27% with the purification function. Finally, the n/gamma resolution and 2D imaging ability of the detector are tested with the Cf-252 neutron source in Institute of High Energy Physics, Chinese Academy of Sciences, the peak ratio of neutron to gamma is obtained to be above 5 from the energy spectra and the detecor has good 2D imaging ability. The performance of the high-pressure MWPC neutron detector could meet the requirements for the multipurpose reflectometer, and the detector will be mounted in the CSNS in this year.
WOS关键词HIGH-PERFORMANCE ; GAS ; DESIGN
WOS研究方向Physics
语种中文
CSCD记录号CSCD:6214973
WOS记录号WOS:000443194200013
内容类型期刊论文
源URL[http://ir.ihep.ac.cn/handle/311005/286255]  
专题高能物理研究所_实验物理中心
高能物理研究所_核技术应用研究中心
中国科学院高能物理研究所_中国散裂中子源
通讯作者祁辉荣
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
刘凌,Sun, ZJ,Li, YH,等. 用于中国散裂中子源多功能反射谱仪的高气压多丝正比室探测器的研制[J]. 物理学报,2018,67(7):72901.
APA 刘凌.,Sun, ZJ.,Li, YH.,Zhang, J.,Wang, YF.,...&温志文.(2018).用于中国散裂中子源多功能反射谱仪的高气压多丝正比室探测器的研制.物理学报,67(7),72901.
MLA 刘凌,et al."用于中国散裂中子源多功能反射谱仪的高气压多丝正比室探测器的研制".物理学报 67.7(2018):72901.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace