Interferences of local B0-shim coils and RF coils on a 3T MRI scanner
Qiaoyan Chen; Jo Lee; Jianghong Wen; Chao Zou; Xiaoliang Zhang; Xin Liu; Ye Li
2018
会议日期2018
会议地点Paris, France
英文摘要In this work, we quantitatively studied the impact of the local shim coil to RF coil in a combined B0 shim coil and RF coil system in terms of SNR, transmit B1 and receive B1. By using the results as a design guideline, a 5-channel shim coil was constructed, of which interferences on RF coils were minimized with the appropriate shim coil diameters, number of turns and distances between the shim coil and the sample.
URL标识查看原文
内容类型会议论文
源URL[http://ir.siat.ac.cn:8080/handle/172644/14549]  
专题深圳先进技术研究院_医工所
推荐引用方式
GB/T 7714
Qiaoyan Chen,Jo Lee,Jianghong Wen,et al. Interferences of local B0-shim coils and RF coils on a 3T MRI scanner[C]. 见:. Paris, France. 2018.