Interferences of local B0-shim coils and RF coils on a 3T MRI scanner | |
Qiaoyan Chen; Jo Lee; Jianghong Wen; Chao Zou; Xiaoliang Zhang; Xin Liu; Ye Li | |
2018 | |
会议日期 | 2018 |
会议地点 | Paris, France |
英文摘要 | In this work, we quantitatively studied the impact of the local shim coil to RF coil in a combined B0 shim coil and RF coil system in terms of SNR, transmit B1 and receive B1. By using the results as a design guideline, a 5-channel shim coil was constructed, of which interferences on RF coils were minimized with the appropriate shim coil diameters, number of turns and distances between the shim coil and the sample. |
URL标识 | 查看原文 |
内容类型 | 会议论文 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/14549] ![]() |
专题 | 深圳先进技术研究院_医工所 |
推荐引用方式 GB/T 7714 | Qiaoyan Chen,Jo Lee,Jianghong Wen,et al. Interferences of local B0-shim coils and RF coils on a 3T MRI scanner[C]. 见:. Paris, France. 2018. |