Fabrication of Dry Electrode for Recording Bio-potentials
Guo K
刊名chinese physics letters
2011
卷号28期号:1页码:article no.10701
关键词AG/AGCL REFERENCE ELECTRODE SENSOR
ISSN号0256-307x
通讯作者wang, y, lanzhou univ, inst microelect, sch phys sci & technol, lanzhou 730000, peoples r china. peiwh@semi.ac.cn
学科主题光电子学
收录类别SCI
资助信息national natural science foundation of china [60776024, 60877035, 90820002]; national high-technology research and development program of china [2007aa03z427, 2007aa04z329, 2007aa04z254]
语种英语
公开日期2011-07-05 ; 2011-07-15
附注development of minimally invasive dry electrodes for recording biopotentials is presented. the detailed fabrication process is outlined. a dry electrode is formed by a number of microneedles. the lengths of the microneedles are about 150 mu m and the diameters are about 50 mu m. the tips of the microneedles are sharp enough to penetrate into the skin. the silver/silver chloride is grown on microneedle arrays and demonstrates good character. the electrocardiogram shows that the dry electrode is suitable for recording biopotentials.
内容类型期刊论文
源URL[http://ir.semi.ac.cn/handle/172111/21021]  
专题半导体研究所_集成光电子学国家重点实验室
推荐引用方式
GB/T 7714
Guo K. Fabrication of Dry Electrode for Recording Bio-potentials[J]. chinese physics letters,2011,28(1):article no.10701.