Vibration Analysis and Optimization of Mobile C-arm CT System
Hongsi Liang; Jing Xiong; Yong Liu; Zhijian Gan; Zeyang Xia
2017
会议日期2017
会议地点美国纽约
英文摘要This paper presents vibration analysis of a mobile C-arm CT system with experimental data. The mobile C-arm CT system under investigation is a device available for three dimensional image reconstruction based on the two dimensional X-ray images generated and relevant sliding motion. This medical device is frequently exposed to vibration in its CT imaging movement, which creates negative impact on the quality of image reconstructed and precision of positioning. In experiment, vibration data is collected and analyzed under C-arm CT working motions. The corresponding new motion planning is designed and tested in experiment. According to the proved experimental data and relevant comparison, the motion planning is optimized for lower vibration amplitude, and it proves the optimized motion planning has a better results compared to the original one. This paper aims to provide suggestions for the motion planning to reduce the mechanical vibration occurred in mobile C-arm CT working condition.
语种英语
内容类型会议论文
源URL[http://ir.siat.ac.cn:8080/handle/172644/12204]  
专题深圳先进技术研究院_医工所
作者单位2017
推荐引用方式
GB/T 7714
Hongsi Liang,Jing Xiong,Yong Liu,et al. Vibration Analysis and Optimization of Mobile C-arm CT System[C]. 见:. 美国纽约. 2017.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace