Investigation of Resonant Frequency and Impedance for Noninvasive Blood Glucose Monitoring
Jingzhen Li; Danlei Yang; Zedong Nie; Yuhang Liu; Lei Wang
2017
会议日期2017
会议地点美国夏威夷
英文摘要Noninvasive blood glucose monitoring (NBGM) provides a prospective approach for patients with diabetes with the advantages of no pain and continuous monitoring. In order to better understand the influence of blood glucose in NBGM, in this paper, the blood mimicking phantoms with different amounts of glucose content were fabricated. Furthermore, the resonant frequency was measured by the vector network analyzer (VNA) when the blood mimicking phantom was attached on the spiral microstrip resonator. In addition, the impedance of blood mimicking phantom was acquired by the impedance analyzer in the frequency range 10 KHz to 1000 KHz when a patch resonator was adopted. The results showed that as the amount of glucose content increased, the resonant frequency increased linearly. However, the change of impedance was nonlinear. The relationship between impedance and the amount of glucose content could be fitted by the quadratic polynomial.
语种英语
内容类型会议论文
源URL[http://ir.siat.ac.cn:8080/handle/172644/12196]  
专题深圳先进技术研究院_医工所
作者单位2017
推荐引用方式
GB/T 7714
Jingzhen Li,Danlei Yang,Zedong Nie,et al. Investigation of Resonant Frequency and Impedance for Noninvasive Blood Glucose Monitoring[C]. 见:. 美国夏威夷. 2017.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace