Investigation of Resonant Frequency and Impedance for Noninvasive Blood Glucose Monitoring | |
Jingzhen Li; Danlei Yang; Zedong Nie; Yuhang Liu; Lei Wang | |
2017 | |
会议日期 | 2017 |
会议地点 | 美国夏威夷 |
英文摘要 | Noninvasive blood glucose monitoring (NBGM) provides a prospective approach for patients with diabetes with the advantages of no pain and continuous monitoring. In order to better understand the influence of blood glucose in NBGM, in this paper, the blood mimicking phantoms with different amounts of glucose content were fabricated. Furthermore, the resonant frequency was measured by the vector network analyzer (VNA) when the blood mimicking phantom was attached on the spiral microstrip resonator. In addition, the impedance of blood mimicking phantom was acquired by the impedance analyzer in the frequency range 10 KHz to 1000 KHz when a patch resonator was adopted. The results showed that as the amount of glucose content increased, the resonant frequency increased linearly. However, the change of impedance was nonlinear. The relationship between impedance and the amount of glucose content could be fitted by the quadratic polynomial. |