a low power adaptive clock and data recovery circuit for wireless implantable systems | |
Yu Hang ; Li Yan ; Jiang Lai ; Ji Zhen ; Yan Ping-Kun ; Wang Fei | |
刊名 | Shenzhen Daxue Xuebao (Ligong Ban)/Journal of Shenzhen University Science and Engineering
![]() |
2011 | |
卷号 | 28期号:2页码:143-146 |
关键词 | Charge pump circuits Clocks CMOS integrated circuits Design Electric power supplies to apparatus Input output programs Integrated circuits Jitter Metallic compounds MOS devices Optical pumping Packet switching Pulse modulation Pulse position modulation Pumps Transistors |
ISSN号 | 1000-2618 |
中文摘要 | An extremely low power clock and data recovery circuit was designed for pulse position modulated input. Synchronized clock and data were recovered through converting the timing distance between pulses into voltage domain. The reference voltage required for data recovery was adaptively generated to extend the range of the input data rate. The design was validated using 0.25 μm CMOS technology. With 45.5 kbits/s input data, the entire circuit only consumes less than 13 μW of power. |
英文摘要 | An extremely low power clock and data recovery circuit was designed for pulse position modulated input. Synchronized clock and data were recovered through converting the timing distance between pulses into voltage domain. The reference voltage required for data recovery was adaptively generated to extend the range of the input data rate. The design was validated using 0.25 μm CMOS technology. With 45.5 kbits/s input data, the entire circuit only consumes less than 13 μW of power. |
收录类别 | EI |
语种 | 英语 |
公开日期 | 2013-10-08 |
内容类型 | 期刊论文 |
源URL | [http://ir.iscas.ac.cn/handle/311060/16175] ![]() |
专题 | 软件研究所_软件所图书馆_期刊论文 |
推荐引用方式 GB/T 7714 | Yu Hang,Li Yan,Jiang Lai,et al. a low power adaptive clock and data recovery circuit for wireless implantable systems[J]. Shenzhen Daxue Xuebao (Ligong Ban)/Journal of Shenzhen University Science and Engineering,2011,28(2):143-146. |
APA | Yu Hang,Li Yan,Jiang Lai,Ji Zhen,Yan Ping-Kun,&Wang Fei.(2011).a low power adaptive clock and data recovery circuit for wireless implantable systems.Shenzhen Daxue Xuebao (Ligong Ban)/Journal of Shenzhen University Science and Engineering,28(2),143-146. |
MLA | Yu Hang,et al."a low power adaptive clock and data recovery circuit for wireless implantable systems".Shenzhen Daxue Xuebao (Ligong Ban)/Journal of Shenzhen University Science and Engineering 28.2(2011):143-146. |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论