CORC  > 清华大学
Sensor factor correction for collimated beam experiments using a medium pressure ultraviolet lamp
Guo, Mei-Ting ; Hu, Hong-Ying ; Bolton, James R.
2010-10-12 ; 2010-10-12
关键词collimated beam ultraviolet UV disinfection radiometer Engineering, Environmental Environmental Sciences
中文摘要A detailed protocol is presented documenting how a 'sensor factor' correction should be calculated when undertaking quasi collimated beam experiments using a polychromatic medium pressure ultraviolet lamp. The correction factor was determined as 0.778 through Pyrex filtration test. This new protocol is particularly important if the radiometer sensor hits significant sensitivity at wavelengths longer than 300 nm.
语种英语 ; 英语
出版者NATL RESEARCH COUNCIL CANADA-N R C RESEARCH PRESS ; OTTAWA ; BUILDING M 55, OTTAWA, ON K1A 0R6, CANADA
内容类型期刊论文
源URL[http://hdl.handle.net/123456789/79872]  
专题清华大学
推荐引用方式
GB/T 7714
Guo, Mei-Ting,Hu, Hong-Ying,Bolton, James R.. Sensor factor correction for collimated beam experiments using a medium pressure ultraviolet lamp[J],2010, 2010.
APA Guo, Mei-Ting,Hu, Hong-Ying,&Bolton, James R..(2010).Sensor factor correction for collimated beam experiments using a medium pressure ultraviolet lamp..
MLA Guo, Mei-Ting,et al."Sensor factor correction for collimated beam experiments using a medium pressure ultraviolet lamp".(2010).