The correlation of the electrode size and S parameter in noninvasive blood glucose detection | |
Xiaoxia Zhang; Yuhang Liu; Zedong Nie; Jingzhen Li; Xiaoling Zhong; Lei Wang | |
2016 | |
会议名称 | ICCC2016会议 |
会议地点 | 中国成都 |
英文摘要 | This paper presents an experimental result towards non-invasive blood glucose detection, which is based on a radiofrequency transmission theory. The experiment is conducted in vitro through 0.0%, 4.0% and 8.0% concentrations of glucose saline solution, respectively. Three different size of square electrodes and round electrodes are adopted in our experiment. For the square electrodes, the sizes of 1.7×1.7cm, 1.3×1.3cm and 1.0×1.0cm are selected. For the round electrodes, the diameters of 2.1cm, 1.7cm and 1.4cm are investigated. The experimental frequency ranges are from 300K to 250MHz, and a VNA is chose for the experimental evaluation. Results show that square and round electrodes with the smallest size have the smallest amplitude of |S11| in the three concentrations of glucose saline solution. |
收录类别 | EI |
语种 | 英语 |
内容类型 | 会议论文 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/10566] |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2016 |
推荐引用方式 GB/T 7714 | Xiaoxia Zhang,Yuhang Liu,Zedong Nie,et al. The correlation of the electrode size and S parameter in noninvasive blood glucose detection[C]. 见:ICCC2016会议. 中国成都. |