A h-shirt-based body sensor network for cuffless calibration and estimation arterial blood pressure
Gu W. B.; Poon C. C. Y.; Sy M. Y.; Leung H. K.; Liang Y. P.; Zhang Y. T.
2009
会议名称Sixth International Workshop on Wearable and Implantable Body Sensor Networks
英文摘要In this paper, a body sensor network (BSN) system for measuring systolic blood pressure (SBP) by a cuffless approach based on h-Shirt has been developed. Two experiments were conducted on a total of 22 subjects to evaluate the cuffless calibration approach and BSN system respectively. The results showed that using the estimated time and distance travelled by a pulse and a parameter derived from the secondary derivative of PPG, the BSN system can measure SBP completely without using a cuff within 1.2±6.0 mmHg of the reference on 10 healthy subjects aged 21-35 yrs.
收录类别EI
语种英语
内容类型会议论文
源URL[http://ir.siat.ac.cn:8080/handle/172644/2566]  
专题深圳先进技术研究院_医工所
作者单位2009
推荐引用方式
GB/T 7714
Gu W. B.,Poon C. C. Y.,Sy M. Y.,et al. A h-shirt-based body sensor network for cuffless calibration and estimation arterial blood pressure[C]. 见:Sixth International Workshop on Wearable and Implantable Body Sensor Networks.