A h-shirt-based body sensor network for cuffless calibration and estimation arterial blood pressure | |
Gu W. B.; Poon C. C. Y.; Sy M. Y.; Leung H. K.; Liang Y. P.; Zhang Y. T. | |
2009 | |
会议名称 | Sixth International Workshop on Wearable and Implantable Body Sensor Networks |
英文摘要 | In this paper, a body sensor network (BSN) system for measuring systolic blood pressure (SBP) by a cuffless approach based on h-Shirt has been developed. Two experiments were conducted on a total of 22 subjects to evaluate the cuffless calibration approach and BSN system respectively. The results showed that using the estimated time and distance travelled by a pulse and a parameter derived from the secondary derivative of PPG, the BSN system can measure SBP completely without using a cuff within 1.2±6.0 mmHg of the reference on 10 healthy subjects aged 21-35 yrs. |
收录类别 | EI |
语种 | 英语 |
内容类型 | 会议论文 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/2566] ![]() |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2009 |
推荐引用方式 GB/T 7714 | Gu W. B.,Poon C. C. Y.,Sy M. Y.,et al. A h-shirt-based body sensor network for cuffless calibration and estimation arterial blood pressure[C]. 见:Sixth International Workshop on Wearable and Implantable Body Sensor Networks. |