Investigation of misalignment in analyzer crystal based-CT and its effect
Zhang K(张凯); Zhu PP(朱佩平); Huang WX(黄万霞); Yuan QX(袁清习); Li ER(李恩荣); Liu YJ(刘宜晋); Liu XS(刘小松); Wang ZL(王志立); Wu ZY(吴自玉); Zhang, K
刊名PHYSICS IN MEDICINE AND BIOLOGY
2008
卷号53期号:20页码:5757-5766
通讯作者[Zhang, Kai ; Zhu, Peiping ; Huang, Wanxia ; Yuan, Qingxi ; Li, Enrong ; Liu, Yijin ; Liu, Xiaosong ; Wang, Zhili ; Wu, Ziyu] Chinese Acad Sci, Inst High Energy Phys, Beijing 100049, Peoples R China ; [Wu, Ziyu] Univ Sci & Technol China, Natl Synchrotron Radiat Lab, Hefei 230026, Peoples R China ; [Wu, Ziyu] Chinese Acad Sci, Theoret Phys Ctr Sci Facil, Beijing 100049, Peoples R China
英文摘要Analyzer crystal-based computed tomography (ACB-CT) is a novel x-ray phase contrast computed tomography technique. In this paper, in order to extract the refraction angle, we analyzed the effect of misalignment on tomography-sliced images of ACB-CT. Two different methods were considered: the first, proposed by Chapman and Dilmanian, was based on two sets of projective image data taken at both sides of the rocking curve and with a half circle sample rotation and the second, recently proposed by our team, was based on only one set of projective image data taken at one side of the rocking curve but with a full circle sample rotation. Theoretical analysis and experimental results will show that the second method improves the quality of reconstructed CT images, also simplifying the ACB-CT data acquisition procedure.
学科主题Engineering; Radiology, Nuclear Medicine & Medical Imaging
类目[WOS]Engineering, Biomedical ; Radiology, Nuclear Medicine & Medical Imaging
研究领域[WOS]Engineering ; Radiology, Nuclear Medicine & Medical Imaging
原文出处SCI
语种英语
WOS记录号WOS:000259736400014
内容类型期刊论文
源URL[http://ir.ihep.ac.cn/handle/311005/240048]  
专题高能物理研究所_多学科研究中心
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhang K,Zhu PP,Huang WX,et al. Investigation of misalignment in analyzer crystal based-CT and its effect[J]. PHYSICS IN MEDICINE AND BIOLOGY,2008,53(20):5757-5766.
APA 张凯.,朱佩平.,黄万霞.,袁清习.,李恩荣.,...&Wu, ZY.(2008).Investigation of misalignment in analyzer crystal based-CT and its effect.PHYSICS IN MEDICINE AND BIOLOGY,53(20),5757-5766.
MLA 张凯,et al."Investigation of misalignment in analyzer crystal based-CT and its effect".PHYSICS IN MEDICINE AND BIOLOGY 53.20(2008):5757-5766.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace